Просмотр исходного кода

Tagit bort många saker som inte kommer hinnas klart. Försökt färdigställa så många kapitel som möjligt. Lagt till ett gäng figurer.

Jonatan Gezelius 5 лет назад
Родитель
Сommit
08c90b8511
30 измененных файлов с 249 добавлено и 171 удалено
  1. 0 0
      diagram/test setup basic.drawio
  2. 0 0
      diagram/test setup pulse calibration.drawio
  3. 0 0
      diagram/test setup pulse injection.drawio
  4. 0 0
      diagram/test setup pulse verification.drawio
  5. 6 0
      rapport/conclusion.tex
  6. 1 0
      rapport/figures/draw/manual-measurement-hv-att-cna.drawio
  7. 1 0
      rapport/figures/draw/manual-measurement-hv-att.drawio
  8. 1 0
      rapport/figures/draw/manual-measurement-hv-diff-cna.drawio
  9. 1 0
      rapport/figures/draw/manual-measurement-hv-diff.drawio
  10. 0 0
      rapport/figures/draw/test setup automatic external.drawio
  11. 0 0
      rapport/figures/draw/test setup automatic internal.drawio
  12. 1 0
      rapport/figures/draw/test setup basic.drawio
  13. 0 0
      rapport/figures/draw/test setup human assisted.drawio
  14. 0 0
      rapport/figures/draw/test setup pulse injection.drawio
  15. 0 0
      rapport/figures/draw/test setup pulse verification.drawio
  16. 0 0
      rapport/figures/draw/testsetup.drawio
  17. BIN
      rapport/figures/manual-measurement-hv-att-cna.png
  18. BIN
      rapport/figures/manual-measurement-hv-att.png
  19. BIN
      rapport/figures/manual-measurement-hv-diff-cna.png
  20. BIN
      rapport/figures/manual-measurement-hv-diff.png
  21. BIN
      rapport/figures/test setup automatic external.png
  22. BIN
      rapport/figures/test setup automatic internal.png
  23. BIN
      rapport/figures/test setup human assisted.png
  24. BIN
      rapport/figures/test setup pulse injection.png
  25. BIN
      rapport/jonatans_exjobb_draft_2020-08-09.pdf
  26. BIN
      rapport/main.pdf
  27. 2 6
      rapport/main.tex
  28. 84 30
      rapport/method.tex
  29. 59 62
      rapport/results.tex
  30. 93 73
      rapport/theory.tex

Разница между файлами не показана из-за своего большого размера
+ 0 - 0
diagram/test setup basic.drawio


Разница между файлами не показана из-за своего большого размера
+ 0 - 0
diagram/test setup pulse calibration.drawio


Разница между файлами не показана из-за своего большого размера
+ 0 - 0
diagram/test setup pulse injection.drawio


Разница между файлами не показана из-за своего большого размера
+ 0 - 0
diagram/test setup pulse verification.drawio


+ 6 - 0
rapport/conclusion.tex

@@ -57,6 +57,12 @@ therein must be concrete and well thought through.
 
 Sätt av ett kort kapitel sist i rapporten till att avrunda och föreslå rikningar för framtida utveckling av arbetet.}
 
+Measurement errors
+
+Mathematial analysis
+
+Automation
+
 %%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
 %%% lorem.tex ends here
 

+ 1 - 0
rapport/figures/draw/manual-measurement-hv-att-cna.drawio

@@ -0,0 +1 @@
+<mxfile modified="2020-08-09T20:28:01.316Z" host="draw.jegatron.se" agent="Mozilla/5.0 (Windows NT 10.0; Win64; x64; rv:79.0) Gecko/20100101 Firefox/79.0" etag="kwrxNeklWHRi76XnqhHh" version="11.2.8" pages="1"><diagram id="RTmf_U5eG0UrcdHPe_QK" name="Page-1">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</diagram></mxfile>

+ 1 - 0
rapport/figures/draw/manual-measurement-hv-att.drawio

@@ -0,0 +1 @@
+<mxfile modified="2020-08-09T20:28:46.367Z" host="draw.jegatron.se" agent="Mozilla/5.0 (Windows NT 10.0; Win64; x64; rv:79.0) Gecko/20100101 Firefox/79.0" etag="9fFUVjuHao3gJlkUixJu" version="11.2.8" pages="1"><diagram id="RTmf_U5eG0UrcdHPe_QK" name="Page-1">1VdNc5swEP01PiaDhMH20SZ2c2gnnXE7aY8qKEACWiqWGPrrKxkBBuyMM3WmiQ9Y+3b19fYtXk9sLy0/SZZFXyDgyYRaQTmxbyaUzi1bPTVQ1cCUujUQyjioIdIB2/gPN6Bl0CIOeN4LRIAE46wP+iAE97GHMSlh1w97gKS/a8ZCPgK2PkvG6H0cYGSu5VgdfsvjMGp2JpbxpKwJNkAesQB2B5C9ntieBMB6lJYeTzR3DS/1vM0Jb3swyQWeM8HLl983z/dErAS59ZyK3Xj5lcnOM0sKc+H15htt7pBj1fAgoRAB10uRib3aRTHybcZ87d2pxCsswjQx7gcQuDVzW7tOrTqkvTKbcom8PHkb0nKktMUh5SgrFWIm2K45o9EVtZ3a3nVZIg310UGGmnnMCCNsl+64UwND3yuoJGPSeKC0ZEyQGEEIgiXrDl0pUWTam8TiSZkdy5qmxyLNDG2uMrsVPgNkhtpHjliZIFYg9BPByxh/HIx/6pWvHWPdlGajvVEZI0cmcalLRwECBG+wTawJqSeIYBChkAN/jhKeuAcJyD0TtrX/tJ6mlE6LQ1P3sjQU01BIn7+UkhMakjxhGD/3NzgmCDP1K8Rq61Z77lB7rtNfQrEVcjSzBrJqj/EPSnPfn9L+t2qOK52+LPVzpXZxBU1ni+u5MyNTWj8XAz3R6wPn3HH769e6H8lrtIszG+h0dp5OVZJYdRCW6YD89G1G+7j2QPb1ihctgvYl8Joi+BCq32zIgiw+vurfVEYfrKQu9uono4ZtichFgQyVfqilG1PrLkrVU9tuojK6+iV7heL+LnTTuVfTVb6X01IFEJqVnVONQv1N9BvYLPlGPSGhl+kJHTpIztQa9YSzIy2h81Yt4XyUrLvcV5UPuQ/qx/i999iufUaPPb1Mj63M7q9QXS3d/0l7/Rc=</diagram></mxfile>

+ 1 - 0
rapport/figures/draw/manual-measurement-hv-diff-cna.drawio

@@ -0,0 +1 @@
+<mxfile modified="2020-08-09T20:19:27.555Z" host="draw.jegatron.se" agent="Mozilla/5.0 (Windows NT 10.0; Win64; x64; rv:79.0) Gecko/20100101 Firefox/79.0" etag="EJZqkh3t49JM8SO82vvg" pages="1" version="11.2.8" type="device"><diagram id="v6DM7zrvzqAhAaCDu8FQ" name="Page-1">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</diagram></mxfile>

+ 1 - 0
rapport/figures/draw/manual-measurement-hv-diff.drawio

@@ -0,0 +1 @@
+<mxfile modified="2020-08-09T20:20:18.372Z" host="draw.jegatron.se" agent="Mozilla/5.0 (Windows NT 10.0; Win64; x64; rv:79.0) Gecko/20100101 Firefox/79.0" etag="4yIYXHlDujktLKxFxptW" pages="1" version="11.2.8" type="device"><diagram id="v6DM7zrvzqAhAaCDu8FQ" name="Page-1">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</diagram></mxfile>

Разница между файлами не показана из-за своего большого размера
+ 0 - 0
rapport/figures/draw/test setup automatic external.drawio


Разница между файлами не показана из-за своего большого размера
+ 0 - 0
rapport/figures/draw/test setup automatic internal.drawio


+ 1 - 0
rapport/figures/draw/test setup basic.drawio

@@ -0,0 +1 @@
+<mxfile modified="2020-08-08T14:47:14.414Z" host="draw.jegatron.se" agent="Mozilla/5.0 (Windows NT 10.0; Win64; x64; rv:79.0) Gecko/20100101 Firefox/79.0" etag="pAEof3MkJ-wOtOPDKcE5" version="11.2.8" type="device" pages="1"><diagram id="v6DM7zrvzqAhAaCDu8FQ" name="Page-1">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</diagram></mxfile>

Разница между файлами не показана из-за своего большого размера
+ 0 - 0
rapport/figures/draw/test setup human assisted.drawio


Разница между файлами не показана из-за своего большого размера
+ 0 - 0
rapport/figures/draw/test setup pulse injection.drawio


Разница между файлами не показана из-за своего большого размера
+ 0 - 0
rapport/figures/draw/test setup pulse verification.drawio


+ 0 - 0
diagram/testsetup.drawio → rapport/figures/draw/testsetup.drawio


BIN
rapport/figures/manual-measurement-hv-att-cna.png


BIN
rapport/figures/manual-measurement-hv-att.png


BIN
rapport/figures/manual-measurement-hv-diff-cna.png


BIN
rapport/figures/manual-measurement-hv-diff.png


BIN
rapport/figures/test setup automatic external.png


BIN
rapport/figures/test setup automatic internal.png


BIN
rapport/figures/test setup human assisted.png


BIN
rapport/figures/test setup pulse injection.png


BIN
rapport/jonatans_exjobb_draft_2020-08-09.pdf


BIN
rapport/main.pdf


+ 2 - 6
rapport/main.tex

@@ -62,12 +62,8 @@
 %%% Non breaking prints of the ISO standards
 \newcommand{\nd}{\nobreakdash-}
 
-%%% Commands for writing out the standards
-\def\setpos#1#2#3{\expandafter\def\csname GO-#1-#2-\endcsname{#3}}
-\def\getpos#1#2{\csname GO-#1-#2-\endcsname}
-
-\setpos{A}{1}{black}
-\setpos{B}{14}{white}
+%% Access to the \tablefootnote command
+\usepackage{tablefootnote}
 
 % Tables something?
 \usepackage{tabularx}

+ 84 - 30
rapport/method.tex

@@ -36,11 +36,7 @@ This chapter covers the methodologies used during the project.
 
 %%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
 \section{Prestudy}
-During the project efforts were made to find relevant research using Linköping University Library's\footnote{\url{https://liu.se/en/library}} and Google Scholar's\footnote{\url{https://scholar.google.se/}} search engines. Among the keywords used in searching were;
-
-\squareit{ \emph{verification equipment}, \emph{test equipment}, \emph{automatic test}, \emph{automatic verification}, \emph{iso equipment}, \emph{electrical verification}, \emph{curve fitting}, \emph{double exponential function}, \emph{ISO 7637}, \emph{ISO 16750}}
-
-\todo[Skriv färdigt nyckelordsdelen och presera dem på ett snyggt men platseffektivt sätt]
+During the project efforts were made to find relevant research using Linköping University Library's\footnote{\url{https://liu.se/en/library}} and Google Scholar's\footnote{\url{https://scholar.google.se/}} search engines.
 
 Since the equipment intended for this project was untested before the project start, the first step was to hook it up and make some initial measurements to be able to decide the continuation of the project.
 
@@ -73,43 +69,86 @@ Since the equipment used in the project is designed for the older version of the
 
 %%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
 \section{Examination and initial measurement of the old equipment}
-To decide the forthcoming of the project, the equipment first had to be checked to see if its performance were within the limits for use with the newer standard. Because there is no dummy loads available at this point of the project, only open load measurements could be done.
-
-With exception for Pulse 3a and Pulse 3b, all of the pulses were measured with the use of the high voltage differential probe described in \autoref{sec:hv-diff-probe}. The pulses are measured both directly on each generator connected according to \autoref{fig:manual-measurement-hv-diff} and also through the coupling network CNA~200, as depicted in \autoref{fig:manual-measurement-hv-diff-cna}. 
-\todo[figure of connection]
-Pulse 3a and Pulse 3b was measured using the attenuators described in \autoref{sec:hv-attenuators} connected according to \autoref{fig:manual-measurement-hv-att}. Thanks to the 50-ohm attenuator this pulse could be measured in its matched state. The measurement in open state is a compromise, since a passive attenuator that does not load the input would be impossible to make, and was made as a 1000-ohm attenuator instead.
+To decide the forthcoming of the project, the equipment first had to be checked to see if it is capable to operate within the limits for use with the newer standard. Because there is no dummy loads available at this point of the project, only open load measurements could be done.
+
+With exception for Pulse 3a and Pulse 3b, all of the pulses were measured with the use of the high voltage differential probe described in \autoref{sec:hv-diff-probe}. The pulses are measured both directly on each generator connected according to \autoref{fig:manual-measurement-hv-diff} and also through the coupling network CNA~200, as depicted in \autoref{fig:manual-measurement-hv-diff-cna}.
+
+\begin{figure}
+	\centering
+	\begin{subfigure}[t]{0.5\textwidth}
+		\includegraphics[width=\textwidth]{manual-measurement-hv-diff}
+		\caption{Without CNA.}
+   	    \label{fig:manual-measurement-hv-diff}
+	\end{subfigure}
+	
+	\begin{subfigure}[t]{0.7\textwidth}
+		\includegraphics[width=\textwidth]{manual-measurement-hv-diff-cna}
+		\caption{With CNA.}
+   	    \label{fig:manual-measurement-hv-diff-cna}
+	\end{subfigure}
+	\caption{The setup for measuring for pulse 1, pulse 2a and Load dump A.}
+\end{figure}
+
+Pulse 3a and Pulse 3b was measured using the attenuators described in \autoref{sec:hv-attenuators} connected directly to the coaxial connector according to \autoref{fig:manual-measurement-hv-att} without the CNA. They were also measured connected through the CNA, directly to the coaxial connector according to \autoref{fig:manual-measurement-hv-att-cna}. Thanks to the 50-ohm attenuator, PAT-50, this pulse could be measured in its matched state. The measurement in open state is a compromise, since a passive attenuator that does not load the input would be impossible to make, and was made as a 1000-ohm attenuator instead.
+
+\begin{figure}
+	\centering
+	\begin{subfigure}[t]{0.45\textwidth}
+		\includegraphics[width=\textwidth]{manual-measurement-hv-att}
+		\caption{Without CNA.}
+   	    \label{fig:manual-measurement-hv-att}
+	\end{subfigure}\hfill
+	\begin{subfigure}[t]{0.45\textwidth}
+		\includegraphics[width=\textwidth]{manual-measurement-hv-att-cna}
+		\caption{With CNA.}
+   	    \label{fig:manual-measurement-hv-att-cna}
+	\end{subfigure}
+	\caption{The setup for measuring for pulse 3a and pulse 3b.}
+\end{figure}
 
 %%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
 \section{Test architecture}
-\squareit{Alternatives and choices. Try finding articles on human error maybe. Make plenty of nice figures.}
 The total number of tests needed to verify the testing equipment before each product test is 14, according to \autoref{tab:verification-list}. There are in total three different values for dummy loads. In practice these could be represented by two high frequency attenuators for pulse 3a and pulse 3b, since these have really short rise times that will be affected much by parasitics of components, and three different high power dummy loads for the slower pulses where the parasitic effects might be negligible but the withstand power must be higher.
 
 The following test architectures were considered, together with the external supervisor at the company.
 
 Additionally there needs to be some sort of measurement fixture for evaluating the verification equipment.
 
+%%%%%%%%%%%%%%%%%%%%%%%%
 \subsection{Alternative 1 -- Human assisted}
-The test can be performed semi-automatically by means of the existing equipment complemented by some dummy loads and, in the same manner the manual performance tests were executed. A computer could control the equipment and compare the results, by the assist of a human that can make the necessary reconnections between the tests.
+The test can be performed semi-automatically by means of the existing equipment complemented by some dummy loads and, in the same manner the manual performance tests were executed. A computer could control the equipment and compare the results, by the assist of a human that can make the necessary reconnections between the tests. A proposed setup for this is shown in \autoref{fig:test_setup_human_assisted}.
 
-\todo[försök hitta källor på följande påståenden]
+The main advantage of this alternative is that it would require the least amount of hardware development time. It also doesn't need any extra hardware except from the dummy loads needed to do the verification.
 
-The main advantage of this alternative is that it would require the least amount of development time. It also doesn't need any extra hardware except from the dummy loads needed to do the verification.
+The biggest disadvantage is that it would be very cumbersome to perform and also prone to human error. If the verification list is studied carefully one can minimize it to five reconnections after the initial connections are made, for example in the following order: No load, \SI{2}{\ohm}, \SI{10}{\ohm}, \SI{50}{\ohm} low frequency, \SI{50}{\ohm} high frequency, \SI{1}{\kilo\ohm} high frequency.
 
-The biggest disadvantage is that it would be very cumbersome to perform and also very prone to human error. If the verification list is studied carefully one can minimize it to five reconnections after the initial connections are made, for example in the following order: No load, \SI{2}{\ohm}, \SI{10}{\ohm}, \SI{50}{\ohm} low frequency, \SI{50}{\ohm} high frequency, \SI{1}{\kilo\ohm} high frequency.
+\begin{figure}[H]
+    %\captionsetup{width=.5\linewidth}
+    \includegraphics[width=0.5\textwidth]{test setup human assisted}    
+    \caption{The proposed setup for alternative must be connected in different ways by a human during the verification process.}
+    \label{fig:test_setup_human_assisted}
+\end{figure}
 
-\subsection{Alternative 2 -- Fully automatic rig external attenuators}
-To accurately measure Pulse 3a and Pulse 3b, the probes should be attached as close as possible to the generator because of the high frequency, to avoid influence of the connecting wires \cite{some_good_reference_for_measurement_techniques} \todo[find source to this]. This could be accomplished by the means of a fixture that is attached directly to the generator, which can switch the pulses to the different loads or to the measurement outputs.
+%%%%%%%%%%%%%%%%%%%%%%%%
+\subsection{Alternative 2 -- Fully automatic rig with external attenuators}
+To accurately measure Pulse 3a and Pulse 3b, the probes should be attached as close as possible to the generator because of the high frequency, to avoid influence of the connecting wires. This could be accomplished by the means of a fixture that is attached directly to the generator, which can switch the pulses to the different loads or to the measurement outputs.
 
-The dummy loads for all pulses, but Pulse 3a and Pulse 3b, will need to be put in a separate enclosure because of the power dissipation needed. The proposed architecture is depicted in \autoref{fig:automatic-rig-1}.
+The dummy loads for all pulses, but Pulse 3a and Pulse 3b, will need to be put in a separate enclosure because of the high power dissipation. The proposed dummy loads for pulse 3a and pulse 3b is the external attenuators PAT~50 and PAT~1000. A proposed setup is depicted in \autoref{fig:test_setup_automatic_external}.
 
-\todo[Fint schema här]
+\begin{figure}[H]
+    %\captionsetup{width=.5\linewidth}
+    \includegraphics[width=0.5\textwidth]{test setup automatic external}
+    \caption{The proposed setup for alternative 2 is fully automatic, but exposes high voltage connectors between the demultiplexer and the two attenuators, marked with a red line.}
+    \label{fig:test_setup_automatic_external}
+\end{figure}
 
 The advantage of this method is that the verification can be performed fully automatically, except for the initial connection of the test rig. This also uses the commercially created attenuators that are already available.
 
 The disadvantage to this setup is that the fixture needs to be designed, making the development costs greater. The fixture that attaches to the generator will expose high voltage on its measurement connectors, making it a safety hazard.
 
+%%%%%%%%%%%%%%%%%%%%%%%%
 \subsection{Alternative 3 -- Fully automatic rig with embedded attenuators}
-To cope with the high voltage exposure, of alternative 1, the high frequency attenuators can be embedded inside the switching fixture, removing the need for high-voltage connectors. \autoref{fig:automatic-rig-2}.
+To cope with the high voltage exposure, of alternative 1, the high frequency attenuators can be embedded inside the switching fixture, removing the need for high-voltage connectors. \autoref{fig:test_setup_automatic_internal}.
 
 To design Alternative 3 some utilities needs to be designed, namely:
 \begin{itemize}
@@ -117,9 +156,14 @@ To design Alternative 3 some utilities needs to be designed, namely:
     \item Match box, the dummy loads with some relays to be able to switch between them.
 \end{itemize}
 
-\todo[Fint schema här]
+\begin{figure}[H]
+    %\captionsetup{width=.5\linewidth}
+    \includegraphics[width=0.5\textwidth]{test setup automatic internal}
+    \caption{The proposed setup for alternative 3 have no high voltage connectors exposed during the calibration.}
+    \label{fig:test_setup_automatic_internal}
+\end{figure}
 
-The advantage of this, on top of the advantages of alternative 2, is that there is no longer need for external attenuators and that the connectors will no longer expose high voltage.
+The advantages of this, in addition to the advantages of alternative 2, are that there is no longer need for external attenuators and that the connectors will no longer expose high voltage.
 
 The disadvantage of this would be that the embedded attenuators might prove difficult to design. They need to be accurate up to high frequencies, be tolerable to high voltage, dissipate the power necessary and also be electrically safe.
 
@@ -129,10 +173,20 @@ Each dummy load must withstand the applied test pulses, and preferably the worst
 
 The dummy loads consists of one or more resistors. When determining whether the resistors withstands the test pulses, the parameters of interest are power dissipation, maximum voltage and maximum energy applied over time.
 
+Since the pulse generators in most cases can generate a higher voltage than required by the standard, the dummy loads should be designed for the worst case setting on the generator. This mitigates the risk of overloading the dummy load caused human error or an error in the control system.
+
+Three different dummy loads are needed. One \SI{2}{\ohm} load for load dump A and for pulse 2a, one \SI{10}{\ohm} load for pulse 1 in \SI{24}{\volt} systems and one \SI{50}{\ohm} load for pulse 1 in \SI{24}{\volt} systems.
+
+%%%%%%%%%%%%%%%%%%%%%%%%
 \subsection{Components}
-At first the momentary worst case powers and voltages were calculated by hand, to the values seen in \autoref{tab:dummy_load_worst_case}. But to find components that can handle these momentary powers proved very difficult, and it is not necessary since the pulse power is only high for a very short time. 
+At first the momentary worst case powers and voltages were calculated by hand, using \autoref{eq:dummy_load_peak}. But to find components that can handle these momentary powers proved very difficult, and it is not necessary since the pulse power is only high for a very short time. 
+
+\begin{equation}
+    P_{peak} = \left( \frac{U_S}{R_S+R_L} \right)^2 R_L
+    \label{eq:dummy_load_peak}
+\end{equation}
 
-One manufacturer of thick film resistors, namely Vishay, specifies its overload capability in a graph with energy over time in the datasheet, which was easier to compare against using LTSpice to simulate the energies for the different loads, according to \autoref{graph:dummy_load_energy}. The simulated value was then divided by the value specified in the datasheet to get the minimum number of resistors required to share the load. Some possible combinations of available resistor values were considered to reach the desired load resistance, before the final configuration were decided according to \autoref{fig:final-dummy-loads}.
+One manufacturer of thick film resistors, namely Vishay, specifies its overload capability in a graph with energy over time in the datasheet. This graph was easier to compare against simulation results achieved with LTSpice, as seen in \autoref{graph:dummy_load_energy}. The simulated value was then divided by the value specified in the datasheet to get the minimum number of resistors required to share the load. Some possible combinations of available resistor values were considered to reach the desired load resistance, before the final configuration were decided according to \autoref{fig:final-dummy-loads}.
 
 \todo[ltspice-bild på de tre olika dummy loadsen]
 
@@ -142,14 +196,14 @@ The voltages used in the calculations are specified in \autoref{tab:dummy_load_w
     \caption{Calculated momentary worst cases for each dummy load.}
 \begin{adjustbox}{width=\columnwidth,center}
     \centering
-    \begin{tabular}{|l|r|r|r|r|r|r|} 
+    \begin{tabular}{|l|r|r|r|r|r|} 
         \hline
-        Dummy load (\si{\ohm}) & Pulse & $R_S$ (\si{\ohm}) & Generator voltage (\si{\volt}) & Peak voltage (\si{\volt}) & Peak power (\si{\watt}) & Mean power (\si{\watt}) \\
+        Dummy load (\si{\ohm}) & Pulse & $R_S$ (\si{\ohm}) & Generator voltage (\si{\volt}) & Resistor voltage (\si{\volt}) & Peak resistor power (\si{\watt}) \\
         \hline
-        2   & Pulse 1 & 2 & 650 & 325 & 45 \si{\kilo} & 5 \\
-        10  & Pulse 1 & 2 & 650 & 600 & 5 \si{\kilo} & 5   \\
-        50  & Pulse 1 & 2 & 650 & 600 & 5 \si{\kilo} & 5   \\
-        50  & Pulse 1 & 2 & 650 & 600 & 5 \si{\kilo} & 5   \\
+        2   & Load dump A & 2 & 200 & 325 & 45 \si{\kilo} \\
+        2   & Pulse 2a    & 2 & 600 & 600 & 5  \si{\kilo} \\
+        10  & Pulse 1     & 2 & 600 & 600 & 5  \si{\kilo} \\ 
+        50  & Pulse 1     & 2 & 600 & 600 & 5  \si{\kilo} \\
         \hline
     \end{tabular}
 \end{adjustbox}

+ 59 - 62
rapport/results.tex

@@ -32,8 +32,9 @@
 
 % !TeX root = main.tex
 \chapter{Results}\label{cha:results}
+This chapter presents the results achieved using the methods described in \autoref{cha:methods}. Each section in this chapter corresponds to a section in the method chapter with the same name.
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%
+%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
 \section{Prestudy}
 Since not much was known about the project at this time, it was difficult to find relevant papers on the topic of the standards.  Most of the literature was found during the project as new problems was found along the way.
 
@@ -43,6 +44,8 @@ The differences of importance between the old and new standards will be presente
 
 One of the most notable differences is the removal of a test pulse from ISO~7637\nd2 that was called \emph{Pulse 5a}. This was instead introduced to the ISO~16750\nd2 under the name \emph{Load dump A}.
 
+Only the properties that proved to differ are mentioned.
+
 %%%%%%%%%%%%%%%%%%%%
 \subsection{Supply voltages}
 The specification of the DC supply voltage for the DUT differs in some case between the older and the newer versions of the standard. There are two different supply voltage definitions. $U_A$ represents a system where the generator is in operation and $U_B$ represents the system without the generator in operation. These have different values for \SI{12}{\volt} and \SI{24}{\volt} systems. $U_B$ is only relevant for Load dump Test A and is thus not defined in ISO~7637 anymore.
@@ -51,7 +54,7 @@ The specification of the DC supply voltage for the DUT differs in some case betw
 
 \begin{table}[H]
     \caption{Comparison of the different supply voltage specifications.}
-\begin{adjustbox}{width=0.8\columnwidth,center}
+\begin{adjustbox}{center}
     %\centering
     \begin{tabular}{|l|r|r|} 
        \hline
@@ -76,44 +79,42 @@ The specification of the DC supply voltage for the DUT differs in some case betw
 
 %%%%%%%%%%%%%%%%%%%%
 \subsection{Surge voltages}
-
-The surge voltages differs\todo[skriv klart!]
-
-Several of the surge voltage ranges has a wider 
+Several of the surge voltages has a wider specified range, as can be seen in \autoref{tab:UADiff}. Notice how the old pulse 5a and the new load dump A have different specifications for $U_S$, but they describe the same pulse because of the different definition of $U_S$ in ISO~7637\nd2 and ISO~16750\nd2.
 
 \begin{table}[H]
     \caption{Comparison of the different surge voltage specifications.}
-\begin{adjustbox}{width=0.8\columnwidth,center}
+\begin{adjustbox}{center}
     %\centering
     \begin{tabular}{|l|r|r|} 
        \hline
         & \multicolumn{2}{c|}{$U_S$}    \\
        Standard & $U_N=$\SI{12}{\volt}  &  $U_N=$\SI{24}{\volt}    \\
        \hline
-       \multicolumn{3}{|c|}{Pulse 1}    \\
+       \multicolumn{3}{|l|}{Pulse 1}    \\
        \hline
        ISO 7637-2:2004   & \SIrange{-75}{-100}{\volt}    & \SIrange{-450}{-600}{\volt}  \\
        ISO 7637-2:2011   & \SIrange{-75}{-150}{\volt}    & \SIrange{-300}{-600}{\volt}  \\
        \hline
-       \multicolumn{3}{|c|}{Pulse 2a}    \\
+       \multicolumn{3}{|l|}{Pulse 2a}    \\
        \hline
-       ISO 7637-2:2004   & \SIrange{37}{50}{\volt}    & \SIrange{37}{50}{\volt}  \\
-       ISO 7637-2:2011   & \SIrange{37}{112}{\volt}   & \SIrange{37}{112}{\volt}  \\
+       ISO 7637-2:2004   & \multicolumn{2}{c|}{\SIrange{37}{50}{\volt}}  \\
+       ISO 7637-2:2011   & \multicolumn{2}{c|}{\SIrange{37}{112}{\volt}}  \\
        \hline
-       \multicolumn{3}{|c|}{Pulse 3a}    \\
+       \multicolumn{3}{|l|}{Pulse 3a}    \\
        \hline
        ISO 7637-2:2004   & \SIrange{-112}{-150}{\volt}    & \SIrange{-150}{-200}{\volt}  \\
        ISO 7637-2:2011   & \SIrange{-112}{-220}{\volt}    & \SIrange{-150}{-300}{\volt}  \\
        \hline
-       \multicolumn{3}{|c|}{Pulse 3b}    \\
+       \multicolumn{3}{|l|}{Pulse 3b}    \\
        \hline
        ISO 7637-2:2004   & \SIrange{75}{100}{\volt}    & \SIrange{150}{200}{\volt}  \\
        ISO 7637-2:2011   & \SIrange{75}{150}{\volt}    & \SIrange{150}{300}{\volt}  \\
        \hline
-       \multicolumn{3}{|c|}{Pulse 5a/Load dump A}    \\
+       \multicolumn{3}{|l|}{Pulse 5a/Load dump A}    \\
        \hline
-       ISO 7637-2:2004   & \SIrange{75}{150}{\volt}    & \SIrange{150}{300}{\volt}  \\
+       ISO 7637-2:2004   & \SIrange{65}{87}{\volt}    & \SIrange{123}{174}{\volt}  \\
        ISO 16750-2:2012 & \SIrange{79}{101}{\volt}    & \SIrange{151}{202}{\volt}   \\
+       ISO 16750-2:2012 \tablefootnote{Recalculated values to fit the same $U_S$ definitions as the older standard. $U_{S_{7637}} = U_{S_{16750}}-U_{N_{16750}}$} & \SIrange{65}{87}{\volt}    & \SIrange{123}{174}{\volt}   \\
        \hline
     \end{tabular}
 \end{adjustbox}
@@ -122,41 +123,11 @@ Several of the surge voltage ranges has a wider
 
 %%%%%%%%%%%%%%%%%%%%
 \subsection{Time constraints}
-Some of the allowed ranges on the pulses parameters has changed compared to the older version of the standard.
-
-\todo[Gör snygga tabeller]
-
-\textbf{Pulse 1}
-Old, 12V: $U_S -75 -- -100$ V
-Old, 24V: $U_S -450 -- -600$ V
-
-New, 12V: $U_S -75 -- -150$ V
-New, 24V: $U_S -300 -- -600$ V
-
-\textbf{Pulse 2a}
-Old: $U_S 37 -- 50$ V
-
-New: $U_S 37 -- 112$ V
-
-\textbf{Pulse 3a and Pulse 3b}
-Old, 12V: $\lvert U_S \rvert 112 -- 150$ V
-Old, 24V: $\lvert U_S \rvert 150 -- 200$ V
-
-New, 12V: $\lvert U_S \rvert 112 -- 220$ V
-New, 24V: $\lvert U_S \rvert 150 -- 300$ V
-
-Old: $t_r 100 -- 200$ µs
-
-New: $t_r 105 -- 195$ µs
-
-\textbf{Load dump A}
-No change, other that the definitions.
-
-
+The only time constraint that is stricter in the newer standard is the risetime of pulse 3a and pulse 3b, $t_r$, as shown in \autoref{tab:timingDiff}
 
 \begin{table}[H]
     \caption{Comparison of the different time constraints.}
-\begin{adjustbox}{width=0.5\columnwidth,center}
+\begin{adjustbox}{center}
     %\centering
     \begin{tabular}{|l|r|} 
        \hline
@@ -172,7 +143,7 @@ No change, other that the definitions.
 \end{table}
 
 %%%%%%%%%%%%%%%%%%%%
-\subsection{Tolerances for verification}
+\subsection{Limits during verification}
 \textbf{Pulse 1}
 
 Old, 24V, matched: $U_S -300 \pm 30$ V
@@ -185,24 +156,50 @@ Old, open: $U_S 50 \pm 5$ V, matched  $U_S 25 \pm 5$ V
 
 New, open: $U_S 75 \pm 7.5$ V, matched  $U_S 35.5 \pm 7.5$ V
 
-\textbf{Pulse 3a and Pulse 3b}
-
-No change.
-
-\textbf{Load dump A}
-
-No change.
-
-%%%%%%%%%%%%%%%%%%%%
-\subsection{Summary of critical changes}
-
-\hl{Fyll i de förändringar som gör sakerna striktare}
-
+\begin{table}[H]
+    \caption{Comparison of the different surge voltage limits during calibration.}
+\begin{adjustbox}{center}
+    %\centering
+    \begin{tabular}{|l|r|r|} 
+       \hline
+        & \multicolumn{2}{c|}{$U_S$}    \\
+       Standard & $U_N=$\SI{12}{\volt}  &  $U_N=$\SI{24}{\volt}    \\
+       \hline
+       \multicolumn{3}{|l|}{Pulse 1}    \\
+       \hline
+       ISO 7637-2:2004   & \SIrange{-75}{-100}{\volt}    & \SIrange{-450}{-600}{\volt}  \\
+       ISO 7637-2:2011   & \SIrange{-75}{-150}{\volt}    & \SIrange{-300}{-600}{\volt}  \\
+       \hline
+       \multicolumn{3}{|l|}{Pulse 2a}    \\
+       \hline
+       ISO 7637-2:2004   & \multicolumn{2}{c|}{\SIrange{37}{50}{\volt}}  \\
+       ISO 7637-2:2011   & \multicolumn{2}{c|}{\SIrange{37}{112}{\volt}}  \\
+       \hline
+       \multicolumn{3}{|l|}{Pulse 3a}    \\
+       \hline
+       ISO 7637-2:2004   & \SIrange{-112}{-150}{\volt}    & \SIrange{-150}{-200}{\volt}  \\
+       ISO 7637-2:2011   & \SIrange{-112}{-220}{\volt}    & \SIrange{-150}{-300}{\volt}  \\
+       \hline
+       \multicolumn{3}{|l|}{Pulse 3b}    \\
+       \hline
+       ISO 7637-2:2004   & \SIrange{75}{100}{\volt}    & \SIrange{150}{200}{\volt}  \\
+       ISO 7637-2:2011   & \SIrange{75}{150}{\volt}    & \SIrange{150}{300}{\volt}  \\
+       \hline
+       \multicolumn{3}{|l|}{Pulse 5a/Load dump A}    \\
+       \hline
+       ISO 7637-2:2004   & \SIrange{65}{87}{\volt}    & \SIrange{123}{174}{\volt}  \\
+       ISO 16750-2:2012 & \SIrange{79}{101}{\volt}    & \SIrange{151}{202}{\volt}   \\
+       ISO 16750-2:2012 \tablefootnote{Recalculated values to fit the same $U_S$ definitions as the older standard. $U_{S_{7637}} = U_{S_{16750}}-U_{N_{16750}}$} & \SIrange{65}{87}{\volt}    & \SIrange{123}{174}{\volt}   \\
+       \hline
+    \end{tabular}
+\end{adjustbox}
+    \label{tab:UADiff}
+\end{table}
 
 %%%%%%%%%%%%%%%%%%%%%%%%%%
 \section{Examination and initial measurement of the old equipment}
 
-\hl{Peta in här någonstans att det var lite krångel att få fart på utrustningen och att det behövdes en del felsökning?}
+At first, the test equipment itself needed some care before it was possible to operate it. A couple of screws were loose inside of the LD~200 and a bridge had to be made for the optional external resistor on the MPG~200 for the pulses to even reach the pulse output connectors.
 
 The result from the initial measurements are presented, along with the limits, in \autoref{tab:initial_measurements} without the CNA~200 connected and in \autoref{tab:initial_measurements_cna} with the CNA~200 connected.
 

+ 93 - 73
rapport/theory.tex

@@ -94,7 +94,7 @@ The general characteristics in common for all pulses are the DC voltage $U_A$, t
 
 An important observation is that the definition of the surge voltage, $U_s$, differs in ISO~7637 and ISO~16750 as depicted in \autoref{fig:loadDumpTestA}.
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
+%%%%%%%%%%%%%%%%%%%
 \subsection{Test pulse 1}
 This pulse simulates the event of the power supply being disconnected while the DUT is connected to other inductive loads. The other inductive loads will generate a voltage transient of reversed polarity onto the DUT's supply lines.
 
@@ -135,7 +135,7 @@ In the standard there are two additional timings associated to this pulse, $t_2$
     \label{tab:pulse1}
 \end{table}
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
+%%%%%%%%%%%%%%%%%%%
 \subsection{Test pulse 2a}
 This pulse simulates the event of a load, parallel to the DUT, being disconnected. The inductance in the wiring harness will then generate a positive voltage transient on the DUT's supply lines. distinguishable when seen on a oscilloscope.
 
@@ -170,7 +170,7 @@ This pulse simulates the event of a load, parallel to the DUT, being disconnecte
     \label{tab:pulse2a}
 \end{table}
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
+%%%%%%%%%%%%%%%%%%%
 \subsection{Test pulse 3a and 3b}
 Test pulse 3a and 3b simulates transients ``which occur as a result of the switching process'' as stated in the standard \cite{iso_7637_2}. The formulation is not very clear, but is interperted and explained by Frazier and Alles \cite{comparison_iso_7637_real_world} to be the result of a mechanical switch breaking an inductive load. These transients are very short, compared to the other pulses, and the repetition time is very short. The pulses are sent in bursts, grouping a number of pulses together and separating groups by a fixed time.
 
@@ -222,7 +222,7 @@ These pulses contain high frequency components, up to 100~MHz, and special care
     \label{tab:pulse3}
 \end{table}
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
+%%%%%%%%%%%%%%%%%%%
 \subsection{Load dump Test A}
 The Load dump Test A simulates the event of disconnecting a battery that is charged by the vehicles alternator, the current that the alternator is driving will give rise to a long voltage transient.
 
@@ -261,7 +261,7 @@ Prior to 2011, the Load dump Test A was part of the ISO~7637-2 standard under th
     \label{tab:loadDumpTestA}
 \end{table}
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
+%%%%%%%%%%%%%%%%%%%
 \subsection{Application of test pulses}
 During a test, the nominal voltage is first applied between the plus and minus terminal of the DUT's power supply input by the test equipment. Then a series of test pulses are applied between the same terminals. The pulses are repeated at specified intervals, $t_1$, as depicted in \autoref{fig:doubleexprep}.
 
@@ -275,7 +275,7 @@ An example of how a test pulse can be applied by the test equipment is depicted
     \label{fig:test_equipment_setup}
 \end{figure}
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
+%%%%%%%%%%%%%%%%%%%
 \subsection{Verification}
 The test pulses are to be verified before they are applied to the DUT. The voltage levels and the timings are to be measured both without any load, and with a matched load, $R_L$ = $R_i$, attached. The standard omits the rise time constraint when the load is attached, except for pulse 3a and 3b. \cite{iso_7637_2}
 
@@ -340,7 +340,7 @@ Chapter 3.1.6 \cite{theCircuitDesignersCompanion}
 \section{Measurement}
 There are several measurement methods needed during the project. To verify the test pulses, voltage has to be measured over time. To verify the dummy loads, resistance has to be measured. To verify the attenuators, their magnitude response has to be measured. This chapter describes the necessary measurement theory required for this project.
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
+%%%%%%%%%%%%%%%%%%%
 \subsection{Resistance}
 To measure resistance, a current is fed through the resistor and the resulting voltage is measured to calculate the resistance using ohms law. This is typically carried out using a multimeter and two probe wires connecting to each terminal of the resistor. When measuring very low valued resistors, however, the resistance in the probe wires can be significant in relation to the resistor measured and will affect the accuracy. One way of overcoming this is to perform a 4-wire measurement using a so called \emph{Kelvin connection}. In this method the current that is fed through the resistor using one pair of wire, and the resulting voltage is measured at the desired point using another pair according to \autoref{fig:kelvin_measurement}.\cite{theCircuitDesignersCompanion}
 
@@ -351,13 +351,13 @@ To measure resistance, a current is fed through the resistor and the resulting v
     \label{fig:kelvin_measurement}
 \end{figure}
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
+%%%%%%%%%%%%%%%%%%%
 \subsection{High Voltage}
 The highest voltage that can be generated by the pulse generators is 1500~V, which is higher than any of the standards require but will serve as the design goal for the verification equipment. This is a higher voltage than most acquisition devices can measure without the use of external attenuators \cite{source}.
 
 Resistive attenuators.. \todo[fyll på]
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
+%%%%%%%%%%%%%%%%%%%
 \subsection{Oscilloscopes, bandwidth, rise time and probes}
 When using an oscilloscope to measure voltage over time, there are several limiting factors to how fast signals one can measure. The oscilloscope itself has a specified bandwidth, as do the probe and any attenuators used. All of these combined determine how short rise times that can be measured accurately. The rise time of the measured will be affected by these properties and the rise time displayed on the oscilloscope screen will be approximately according to \autoref{equ:riseComposite}, where $T_N$ is the \SIrange{10}{90}{\percent} rise time limit for each part in the chain. \cite{highSpeedDigitalDesign}
 
@@ -373,15 +373,11 @@ Since \autoref{equ:riseComposite} is based on the rise time limitation but the s
 T_{10-90} = \frac{0.338}{F_{ \SI{3}{\deci\bel}}}
 \end{equation}
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
-\subsection{Measurement errors}
-\todo[Put good theory here]
-
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
+%%%%%%%%%%%%%%%%%%%
 \subsection{RF Attenuators}
 Linearity, tolerances, power, combinations of resistors, impedances
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
+%%%%%%%%%%%%%%%%%%%
 \subsection{Tolerances and maximum ratings}
 Resistors, Power, Voltages, surges,
 Relays, isolation, dielectric strength
@@ -390,7 +386,7 @@ Relays, isolation, dielectric strength
 \section{Analysis}
 The data points from the measurement must be processed and evaluated to determine if the measured pulse is within the specified limits.
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
+%%%%%%%%%%%%%%%%%%%
 \subsection{Mathematical description}
 All of the test pulses applied to the vehicle equipment can individually be described mathematically by variations of the double exponential function shown in \autoref{eq:doubleexp} and \autoref{fig:doubleexp}. The properties of interest, the ones which are specified in the standards, are the surge voltage $ U_s $, the rise time $ t_r $, the duration $ t_d $ and the repetition time $ t_1 $. \cite{iso_7637_2}
 
@@ -399,28 +395,7 @@ All of the test pulses applied to the vehicle equipment can individually be desc
     \label{eq:doubleexp}
 \end{equation}
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
-\subsection{What is good}
-\label{sec:goodness}
-\todo[Någonting om vad som anses bra]
-
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
-\subsection{Curve fitting?}
-\todo[Läs på om ämnet och se ifall det kan vara rimligt]
-
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
-\subsection{Max/min limits?}
-\todo[Användandet av max/min-fönster]
-
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
-\subsection{Parameter extraction?}
-\todo[Detta är nog ett påhittat ord, kanske menar jag curve fitting?]
-
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
-\subsection{Evaluation/simulation/robustness}
-\todo[Jämför och evaluera de två eller tre metoderna med hänseende till vad som står i ``goodness'']
-
-----
+It is not in the scope of this report to actually fit this function to the measured pulse, and further analyze it.
 
 \squareit{
 Essentially an embedded high voltage probe.
@@ -475,61 +450,106 @@ On the  Design  and  Generation of the Double  Exponential Function S. C. Dutta
 \section{Instrumentation and control}
 The following chapter describes the different instruments that were used, and their control interfaces.
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
+%%%%%%%%%%%%%%%%%%%
 \subsection{GPIB}
-IEEE-488, or GPIB which it is often called, is a parallel bus interface. It is mainly used to interconnect lab instrumentation such as multimeters, signal generators and spectrum analyzers. 
-\todo[fyll på och hitta källor, lägg in bild på interface]
+IEEE-488, or GPIB which it is often called, is a parallel bus interface. It is mainly used to interconnect lab instrumentation such as multimeters, signal generators and spectrum analyzers.
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
+%%%%%%%%%%%%%%%%%%%
 \subsection{Tektronix TDS7104 Oscilloscope}
-The oscilloscope that is available is a Tektronix TDS7104, with specifications as seen in \autoref{tab:tds7104}. It has GPIB interface and TekVISA GPIB, an API for sending GPIB commands over ethernet, available for remote control. 
+The oscilloscope that is available is a Tektronix TDS7104, with specifications as seen in \autoref{tab:tds7104}. It has GPIB interface and TekVISA GPIB, an API for sending GPIB commands over ethernet, available for remote control.
 
-\todo[Lägg in bild på utrustning, och tabell med data]
+%%%%%%%%%%%%%%%%%%%
+\subsection{xxxxx Isolated differential probe}
+\label{sec:hv-diff-probe}
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
+%%%%%%%%%%%%%%%%%%%
 \subsection{EM Test MPG 200 Micropulse generator}
-The MPG~200 is used to generate \emph{Test pulse 1} and \emph{2a}. MPG is an abbreviation for \emph{MicroPulse Generator}. The instrument is designed to generate test pulses according to the older ISO~7637-2:1990 version, but the parameters can be adjusted to comply with the new ISO~7637:1990 standard.
+The MPG~200 is used to generate \emph{Test pulse 1} and \emph{2a}. MPG is an abbreviation for \emph{MicroPulse Generator}. The instrument is designed to generate test pulses according to the older ISO~7637-2:1990 version, but the parameters can be adjusted to comply with the new ISO~7637:1990 standard. The adjustable parameter ranges are shown in \autoref{tab:mpg200_specs}.
 
-\todo[Lägg in bild på utrustning, och tabell med data]
+\begin{table}[H]
+    \caption{Adjustable parameters in the MPG 200}
+\begin{adjustbox}{center}
+    %\centering
+    \begin{tabular}{|l|r|} 
+        \hline
+        Parameter & Range \\
+        \hline
+        $U_S$  & \SIrange{20}{600}{\volt} \\
+        $U_S$ polarity & +, - \\
+        $R_s$  & \SIlist{2;4;10;20;30;50}{\ohm}  \\
+        $t_1$  & \SIrange{0.2}{99.0}{\second} \\
+        $t_2$  & \SIrange{0}{10}{\second} \\
+        \hline
+    \end{tabular}
+\end{adjustbox}
+    \label{tab:mpg200_specs}
+\end{table}
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
+%%%%%%%%%%%%%%%%%%%
 \subsection{EM Test EFT 200 Burst generator}
-The EFT~200 is used to generate \emph{Test pulse 3a} and \emph{3b}. EFT is an abbreviation for \emph{Electrical Fast Transient}. The instrument is designed to generate test pulses according to the older ISO~7637-2:1990 version, but the parameters can be adjusted to comply with the new ISO~7637:1990 standard.
+The EFT~200 is used to generate \emph{Test pulse 3a} and \emph{3b}. EFT is an abbreviation for \emph{Electrical Fast Transient}. The instrument is designed to generate test pulses according to the older ISO~7637-2:1990 version, but the parameters can be adjusted to comply with the new ISO~7637:1990 standard. The adjustable parameter ranges are shown in \autoref{tab:eft200_specs}.
 
-\todo[Lägg in bild på utrustning, och tabell med data]
+\begin{table}[H]
+    \caption{Adjustable parameters in the EFT 200}
+\begin{adjustbox}{center}
+    %\centering
+    \begin{tabular}{|l|r|} 
+        \hline
+        Parameter & Range \\
+        \hline
+        $U_S$  & \SIrange{25}{1500}{\volt} \\
+        $U_S$ polarity & +, - \\
+        Coupling & any combination of +, - and GND \\
+        \hline
+    \end{tabular}
+\end{adjustbox}
+    \label{tab:eft200_specs}
+\end{table}
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
+%%%%%%%%%%%%%%%%%%%
 \subsection{EM Test LD 200 Load dump}
-The LD~200 is used to generate \emph{Load dump Test A}. LD is an abbreviation for \emph{Load Dump}. The instrument is designed to generate test pulses according to the older ISO~7637-2:1990 version, but the parameters can be adjusted to comply with the new ISO~16750:2012 standard.
+The LD~200 is used to generate \emph{Load dump Test A}. LD is an abbreviation for \emph{Load Dump}. The instrument is designed to generate test pulses according to the older ISO~7637-2:1990 version, but the parameters can be adjusted to comply with the new ISO~16750:2012 standard. The adjustable parameter ranges are shown in \autoref{tab:ld200_specs}.
 
-\todo[Lägg in bild på utrustning, och tabell med data]
+\begin{table}[H]
+    \caption{Adjustable parameters in the LD 200}
+\begin{adjustbox}{center}
+    %\centering
+    \begin{tabular}{|l|r|} 
+        \hline
+        Parameter & Range \\
+        \hline
+        $U_S$  & \SIrange{20}{200}{\volt} \\
+        $R_s$  & \SIlist{0.5;1;2;10}{\ohm}  \\
+        $t_d$  & \SIrange{50}{400}{\milli\second} \\
+        \hline
+    \end{tabular}
+\end{adjustbox}
+    \label{tab:ld200_specs}
+\end{table}
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
+%%%%%%%%%%%%%%%%%%%
 \subsection{EM Test CNA 200 Coupling Network}
-The SNA~200 is a coupling network, used to multiplex the pulse generators into one box. It contains several relays to select the appropriate generator output. The SNA~200 has one interface for each pulse generator, but no interface for a computer. It is automatically controlled by the pulse generators.
+The SNA~200 is a coupling network used to multiplex the pulse generators outputs. It contains several relays to select the appropriate generator output. The SNA~200 has one interface for each pulse generator, but no interface for a computer. It is automatically controlled by the pulse generators.
 
-\todo[Lägg in bild på utrustning, och tabell med data]
+This allows the DUT to be connected only to the SNA~200 and not to each individual pulse generator. \autoref{fig:test_setup_cna_dut} shows the connections between the instruments in this setup.
 
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
+There is also a coaxial connection for calibration of pulse 3a and pulse 3b on the front panel.
+
+\begin{figure}[H]
+    %\captionsetup{width=.5\linewidth}
+    \includegraphics[width=0.5\textwidth]{test setup pulse injection}    
+    \caption{The CNA~200 allows each pusle generator to output their pulses through a common interface towards the DUT.}
+    \label{fig:test_setup_cna_dut}
+\end{figure}
+
+
+%%%%%%%%%%%%%%%%%%%
 \subsection{Rohde \& Schwarz ZVL13}
 \label{sec:rohde_schwarz_zvl}
 The ZVL13 is a vector network analyzer. It is, in this project, used to measure the magnitude and phase response between its two ports.
 
-\todo[Lägg in bild på utrustning, och tabell med data]
-
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
+%%%%%%%%%%%%%%%%%%%
 \subsection{PAT 50 and PAT 1000}
-\label{theory_pat_attenuators}
-
-
-
-
-
-
-%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
-%%% lorem.tex ends here
+\label{sec:hv-attenuators}
+\todo[Skriv någonting här!]
 
-%%% Local Variables: 
-%%% mode: latex
-%%% TeX-master: "demothesis"
-%%% End: 

Некоторые файлы не были показаны из-за большого количества измененных файлов